SC Source Series Data Sheet
SC005 & SC008 – constant current / constant voltage types

FAST POLARITY SWITCHING HIGH VOLTAGE MODULES

Applications:

Mass spectrometer Electro Spray Ionization & Atmospheric Pressure Chemical Ionization source

  • Constant voltage ESI: ±5kV or ±8kV types @ 500µA / 300µA & 2µA for nano spray applications
  • Tru-Current™ APCI: <2μA to ±50μA with adjustable compliance voltage
  • Fast polarity switching within 10mS or 50mS types
  • True zero volt crossing
  • Short circuit and flashover proof, fully screened metal case for good EMI immunity

This range of high performance mass spectrometer source supplies has been specifically designed to meet the needs of today’s high throughput mass spectrometers. Featuring Tru-Current™ technology with control at the APCI source, and fast polarity changeover in less than 10mS, these supplies are ideal for multimode source applications where changeover and setting time are critical.

Electrical Specification

Unit Type

Mode(s)
Output voltage & current range
Changeover time
SC005RCV010 Constant voltage 0 to ±5kV at ±500µA <10mS
SC008RCV010 Constant voltage 0 to ±8kV at ±300µA <10mS
SC008RCV010-1x Constant voltage 0 to ±8kV at ±2µA (Nano amp current monitor) <10mS
SC008RCC010 Constant current 0 to ±50µA at ±8kV compliance (variable) <10mS
SC005RCD010 Constant current & voltage 0 to ±50µA / ±500µA at 0 to ±5kV <10mS
SC008RCD010 Constant current & voltage 0 to±50µA /±300µA at 0 to ±8kV <10mS
SC00xRCx050
-
As above for 50mS equivalents <50mS


For corona source applications (constant current mode) the minimum controllable current is 2μA. For
special versions of these units we would be pleased to discuss your application with you.

Input:

+24 volt dc ±10% <1A. 0V input common to HV return and chassis.

Output, voltage / current

See table above.

Ripple: <0.1% p/p
Line regulation: <300ppm of rated output for ±10% input change.
Load regulation: <0.5% at rated output.
Temperature coefficient: <100ppm/oC at rated output
Drift (after 1 hour warm up): 100ppm per hour, 200ppm per 8 hours.
Control input:
V for CV option, C for CC option
0 to +5V analogue input for rated negative or positive output, accuracy ±2%,
Zin 100K.
Compliance input:
C for CV option, V for CC option
0 to +5V analogue input for rated output, accuracy ±2%, Zin 100K.
User accessible preset potentiometer, 0 to +5V available (pin 3).
Mode input (RCD types): TTL: Hi = constant voltage, Lo = constant current, internal +5V pull up.
Voltage & Current monitors: 0 to +5V for rated negative or positive output, accuracy ±2%, Zout 10K.
Polarity input TTL: Hi = +ve, Lo = -ve, internal +5V pull up.
Safety Conforms to EN61010-1:2001 (stored charge is <45uC).


Mechanical Specification

Dimensions:

146 x 133 x 48 mm
(5.75” x 5.25” x 1.9”)

Weight:

1200g (2.6 lb) approx

Mountings: 6 off blind M3 bushes
Input & control 15 way “D” connector
Output (CV only types)
(CC types)
0.75m of URM 76 cable
0.75m of Belden cable
(all are un-terminated, or customer specified connector to order)

PDF Downloads

Data Sheet
Home | Company Profile | M/S Applications | Contact Us | Representatives