HC Series Data Sheet

HC010R, HC020R and HC030R
100W
FAST POLARITY SWITCHING MODULES

Application

MALDI Mass spectrometers, Nuclear instrumentation,
capillary electrophoresis

  • ±10kV, ±20kV and ±30kV types
  • Fast solid state polarity switching
  • Remote voltage programming
  • High Stability, temp-co < 25ppm
  • Voltage monitor & current monitor
  • Flashover & short circuit protected

This range of high performance 100W high voltage power supplies has been specifically designed for applications where a remotely switchable polarity is required. This compact converter module employs solid state switching techniques allowing for fast polarity reversal whilst maintaining excellent line and load regulation. These modules are designed to operate from a dc supply of 24V (nominal) and provide an accurately controlled high voltage output. The use of high stability components within the feedback system ensures a low temperature coefficient and good long & short term drift making the HC series ideal for use in MALDI time of flight mass spectrometry equipment with pulsed load requirements. The output voltage may be controlled from 50V to maximum (10, 20 or 30kV) of either polarity by means of an external control voltage (0 to +10V). Monitoring of output voltage, current and polarity status are provided. Current sink & source versions are also available.


Electrical Specification

Unit Type

Output Voltage

Output Current

Ripple (pk-pk)

Temp-Co ( / degC)
HC010RAA050 ±10kV 10mA 100ppm 50ppm
HC020RAA050
±20kV
5mA
100ppm
50ppm
HC030RAA050
±30kV
3.3mA
100ppm
50ppm
HC0XXRAA025
As above for 25ppm equivalents
25ppm
HC0XXRPR0XX
As above with potentiometer & reference fitted

Input: +24 volt dc ±10% <6A. 0V input common to HV return and chassis
Output, voltage / current / ripple: See table above.
Line regulation: <10ppm for a 1V change in input voltage.
Load regulation: <10ppm for zero to full load.
Drift (after 1 hour warm up): <100ppm per hour, <200ppm per 8 hours
Protection: Current limit to 110% (typ) auto recovery. Protected against intermittent flashover to ground
Control of output: True differential analogue input of zero to 10V dc varies output from zero to maximum
+Ve input R=100k, input range: -1V to +10V, with respect to ground
-Ve input R=100k, input range: +1V to -10V, with respect to ground
Output slew rate limit:

Linear ramp 500mS / 10kV (faster rate available to special order)

Polarity control: Negative = High (>2.5V) or open circuit. Positive = Low (<0.8V)
Polarity change time: <250mS. Between maximum rated output to maximum rated output of the opposite polarity
Voltage monitor: Zero to +10V ±2% for zero to maximum (Z= 10k)
Current monitor: Zero to +10V ±2% for zero to maximum (Z= 10k)

Mechanical Specification

Dimensions:

240 x 216 x 85 mm (9.45” x 8.5” x 3.35”)

Weight: 3.8kg (8.3 lb) approximately
Mountings centres: 230.3 x 175.6 mm (9.067” x 6.9”) 4 off M4 bushes
Input / control: 12 way 0.2” Molex connector
Output Flying lead, red UL3239 screened 30kV cable, 1m long

PDF Downloads

Data Sheet

Home | Company Profile | M/S Applications | Contact Us | Representatives