3.5kV Floating Microchannel Plate Power Supplies for Mass Spectrometers at Up to 30kV Isolation

Applications:

  • Microchannel plates for
    • Mass Spectrometers
    • Electron Microscopes
  • Floating grid voltages
  • Floating bias voltages
  • NEW Low Injected Ripple Version

These power supplies, based on the standard HP series are intended to power imaging detectors. They give 0 to +3.5kV and can be floated on voltages up to ±2.5kV (HP3F2.5), ±10kV (HP3F10), ±20kV (HP3F20) & ±30kV (HP3F30) and use differential feedback techniques to allow the 24 volt power, control and monitor signals to be at ground potential.

Change to HP3F series;
The Internal Potentiometer and Reference, necessary when using internal or external potentiometer Control, are now only fitted as an optional extra. Please request 'Plus Ref' on order.

Ordering information: State unit required e.g. HP3F10L (3.5 kV output floating up to 10 kV with low injected ripple option)

Electrical Specification

UNIT TYPE

OUTPUT

ISOLATION

OUTPUT RIPPLE INJECTED RIPPLE (1)

NEW Low Injected Ripple Option

HP3F2.5(L) 0+3.5kV
at 1 ma
±2.5kV (2) 50mV peak to peak 50mV peak to peak 25mV - L option
HP3F10(L) 0+3.5kV
at 1 ma
±10kV (3) 75mV peak to peak 75mV peak to peak 35mV - L option
HP3F20(L) 0+3.5kV
at 1 ma
±20kV (3) 100mV peak to peak 150mV peak to peak 75mV - L option
HP3F30 0+3.5kV
at 1 ma
±30kV (3) 150mV peak to peak 200mV peak to peak -

(1) ripple injected into the power supply providing the floating voltage, measured assuming load capacitance of 1000°F.

(2) resistance to ground 400 Mohm on each output.

(3) resistance to ground 600 Mohm on each output.

INPUT VOLTAGE:

+24 volt at 0.7 amp at 20 mA
(HP3F10 & HP3F20)..

OUTPUT VOLTAGE:

See table above.

CONTROL:

10V analog signal. (0 to +10V gives zero to maximum output, tolerance ±3%). Input impedance > 100Kohm. INTERNAL or EXTERNAL potentiometer. Option - quote +Ref with Order.

LINE REGULATION:

Less than 0.1%.

LOAD REGULATION:

Less than 0.1%.

VOLTAGE MONITOR:

0 to +10V ±3% represents 0 to +3.5kV.
Source resistance 10 kohm.
DRIFT: Typically 0.1% of full scale voltage per hour, measured at constant input voltage, output voltage, load, polarity & ambient temperature after 1 hour's warm up.

TEMPERATURE COEFFICIENT:

Less than 0.02% per °C.

OPERATING TEMPERATURE:

0°C to +50°C.

Mechanical Specification

Size:

200 x 98 x 47mm (standard HP10P case).
HP3F10, HP3F20 & HP3F30: 210 x 120 x 60mm.
INPUT: Molex connector. (24V power, control and monitor are all at ground potential).
FLOATING OUTPUT: By twin flying screened (shielded) leads type URM43 (HP3F2.5 & HP3F10), TV30 (HP3F20/30).