Reversible High Voltage Modules + Floating 3kv Detector Supply

Applications:
  • TOF Mass Spectrometers
  • Floating Detectors

NEW! RD now extended to 12.5Kv

Pin Connection & Safety Instructions

Electrical Specification

UNIT TYPE

OUTPUT

RIPPLE AT FULL LOAD *

FLOATING DETECTOR

RD005RIP025 -5 kV to +5 kV at 250uA 150mV peak to peak 3Kv @ 400µA <200mV Ripple
RD7.5RIP025 -7.5kV to +7.5kV at 150µA 165mV peak to peak 3Kv @ 400µA <200mV Ripple
RD010RIP025 -10kV to +10kV at 125µA 200mV peak to peak 3Kv @ 400µA <200mV Ripple
RD012RIP025 -12.5kV to +12.5kV at 150µA 200mV peak to peak 3Kv @ 400µA <200mV Ripple

INPUT VOLTAGE:

+24V ±10% at 1 A.

BIAS OUTPUT VOLTAGE:

See table above. Joined within the unit to Detector -ve

BIAS CONTROL VOLTAGE:

0V to +10V gives 0V to max. kV to ±2%.
nb: Internal manual potentiometer control not fitted.
LOAD REGULATION: Less than 100 ppm for zero to max. load.

LINE REGULATION:

Less than 100 ppm for 1V change in I/P volts.
V MONITOR 0 to ±10V represents 0 to ±max o/p voltage, ±2%. output impedance 10 kohm.

I MONITOR:

0 to ±10V represents 0 to ±max o/p current, ±5%. output impedance 10 kohm.
SPEED OF REVERSING Better than 500 mS.

TEMPERATURE COEFFICIENT:

<25 ppm per °C. (<10ppm per °C available shortly)

DETECTOR CONTROL VOLTAGE:

0V to +10V gives 0V to 3kV ±5%.

MINIMUM DETECTOR O/P V:

500V

DETECTOR VMON:

0 to 10V represents 0 to 3kV o/p voltage, output impedance 10 kohm.
DETECTOR LOAD REGULATION:

Less than 0.1% for zero to max. load.

DETECTOR LINE REGULATION:

Less than 0.1% for 1V change in I/P volts.

DETECTOR TEMP-CO <300 ppm per °C.

OPERATING TEMPERATURE:

0°C to +45°C.

Mechanical Specification

SIZE:

240 x 216 x 100mm. NB. Grounding of the case is via fixing bolts.

OUTPUT CABLE:

URM43 length 0.5 metre. Labelled Detector Minus, & Detector Plus
INPUT CONNECTOR: BERG 20 Way IDC header.
COMPLIANCE

Designed to meet or exceed EN61010.