Excellence in High Voltage
LS Series Data Sheet

LS001, LS2.5, LS005, LS010,
LS015, LS020, LS030, LS050,

PRECISION HIGH VOLTAGE MODULES

Applications (1 Watt max)
See LT Series for 3 Watt

  • Mass spectrometers
  • Electron microscopes
  • Nuclear instruments
  • Photomultiplier tubes
  • Microchannel plates



Features

  • 1kV, 2.5kV, 5kV, 10kV, 15kV, 20kV, 30kV, 50kV
  • Highest Stability, Lowest Ripple, Lowest Drift
  • Externally programmable or Internal control (option)
  • Short circuit and flashover proof
  • 24 hour burn in
  • Low radiated magnetic field
  • Positive, negative & remote reversible versions

The LS Series sets a new standard for high voltage precision and stability. By optimising the Low Stress™ technologies used in our leading 10 Watt HP Series for 1 Watt applications, the new LS Series has less than 1ppm ripple, 10ppm/°C temperature coefficient, and 50ppm drift/hr throughout the range. Each of the units includes a differential control input. For 3W versions see our LT Famly.

To take full advantage of the performance available from the LS Range, they are initially only being made available to our OEM customers as application specific, custom optimised, modules.

All units are short circuit proof and use proven techniques to drive a high frequency oscillator and ferrite high voltage step-up transformer. These advanced power supplies build on Applied Kilovolts’ legendary reputation for reliability, being constructed from conservatively rated components with their reliability further enhanced by information gained over many years of field operation.

Available as positive or negative up to 50kV, and electrically reversible at voltages up to 30kV.

Please see the HPRZC range for zero crossing reversible (bi-polar) units.


Electrical Specification
Positive & Negative Polarity Units

UNIT TYPE POLARITY OUTPUT

RIPPLE AT FULL LOAD

SIZE (mm)
LS001PIP010

POSITIVE

10 volts to 1kV at 1mA

2mV peak to peak

156x127x64

LS001NIP010

NEGATIVE

LS2.5PIP010

POSITIVE

10 volts to 2.5kV at 400uA

3mV peak to peak

156x127x64

LS2.5NIP010

NEGATIVE

LS005PIP010

POSITIVE

10 volts to 5kV at 200uA

5mV peak to peak

156x127x64

LS005NIP010

NEGATIVE

LS010PIP010

POSITIVE

20 volts to 10kV at 100uA

10mV peak to peak

156x127x64

LS010NIP010

NEGATIVE

LS015PIP010

POSITIVE

30 volts to 15kV at 66.7uA

15mV peak to peak

207x148x74

LS015NIP010

NEGATIVE

LS020PIP010

POSITIVE

50 volts to 20kV at 50uA

20mV peak to peak

207x148x74

LS020NIP010

NEGATIVE

LS030PIP010

POSITIVE

100 volts to 30kV at 33.3uA

30mV peak to peak

207x148x74

LS030NIP010

NEGATIVE

LS050PIP010

POSITIVE

200 volts to 50kV at 20uA

50mV peak to peak

TBD

LS050NIP010

NEGATIVE


Electrical Specification
Reversible Polarity Units

UNIT TYPE

POLARITY

OUTPUT

RIPPLE AT FULL LOAD

SIZE (mm)

LS001RIP010

REVERSIBLE

±10 volts to ±1kV at 1mA

2mV peak to peak

163x152x71.5

LS2.5RIP010

REVERSIBLE

±10 volts to ±2.5kV at 400uA

3mV peak to peak

163x152x71.5

LS005RIP010

REVERSIBLE

±10 volts to ±5kV at 200uA

5mV peak to peak

163x152x71.5

LS010RIP010

REVERSIBLE

±20 volts to ±10kV at 100uA

10mV peak to peak

163x152x71.5

LS015RIP010

REVERSIBLE

±30 volts to ±15kV at 66.7uA

15mV peak to peak

233x204x114

LS020RIP010

REVERSIBLE

±50 volts to ±20kV at 50uA

20mV peak to peak

233x204x114

LS030RIP010

REVERSIBLE

±100 volts to ±30kV at 33.3uA

30mV peak to peak

233x204x114

See HPZ data sheet for continuously variable zero crossing power supplies


INPUT VOLTAGE: +24 volt d.c. ±10% at less than 0.3A. Negative input terminal common to HV earth return.
OUTPUT VOLTAGE: See tables above.
LINE REGULATION: Less than 2ppm of maximum voltage for input changes of 1 volt.
LOAD REGULATION: Less than 10ppm for load changes from 10% to 100%. (Measured at maximum voltage).
OVERLOAD PROTECTION: Flashover and short circuit proof. Re-entrant current limit.
TEMPERATURE COEFFICIENT: <10ppm/°C
DRIFT:
<50ppm per hour (<0.005%)
<100ppm per 8 hours (<0.01%)
Measured at constant input voltage, load current & ambient temperature after 0.5 hour warm up.
CONTROL: 10V analogue signal. (0 to +10V gives zero to maximum output, tolerance ±2%). Input impedance > 100Kohm.
EXTERNAL potentiometer = Option Code PP INTERNAL pote ntiometer = Option Code FP
READOUT: Voltage monitor: 0 to +10V represents zero to maximum output, tolerance ±2% Zout LS=10k, Zout LSR=0k
Precision Current Monitor: 0 to 10V represents zero to 100% o/p current, tolerance ±2%, Offset ±0.1% of FS.Zout 10k
OPERATING TEMPERATURE: 0°C to +50°C
STORAGE TEMPERATURE: 35°C to +85°C
R.F.I.: Steel case for low radiated magnetic field.

Mechanical Specification

FIXINGS:

M3 Blind Fasteners front and rear. (10mm deep)

OUTPUT:

By 1 metre screened cable, or optional integrated high voltage connector N.B. Reducing the cable length may increase the ripple voltage.
INPUT CONNECTOR: 15 way D-type
e.g. Order Code: LS010NIP010 = negative 10kV unit LS005PPP010= +ve 5kV unit for control by External Pot

PDF Downloads

Data Sheet

Home | Company Profile | M/S Applications | Contact Us | Representatives
Copyright © Applied Kilovolts Ltd. 2004